AirSep Corp. — FreeStyle Service Manual

5-23

9. Remove hour meter as per Section 5.12.

10. Remove cannula nozzle tubing by cutting tie-wrap (use small-size cutter).

Carefully pull tubing from nozzle barb. (Figure 5.32)

Figure 5.32

11. Disconnect fan connector using needle-nose pliers, carefully pulling upward

on fan connector housing at location J6 on main circuit board.

Figure 5.33

12. Remove six 4-20 self-tapping plastic screws.

(Figure 5.33)