beautypg.com

4 freestyle specifications – AirSep MN126-1 C User Manual

Page 10

background image

AirSep Corp. — FreeStyle Service Manual

1-4

1.4 FreeStyle Specifications










Max Operational
Altitude:



up to 12,000 ft (3,657.6 m) (483mmHg) Higher altitudes may affect
performance

* Based on an atmospheric pressure of 14.7 psia (101 kPa) at 70

o

F (21

o

C)

* Operating unit outside of normal operational temperature range can affect performance.

Oxygen Concentration:*

1-3 pulse settings; equivalent to a continuous flow of

90% +5.5/-3% Oxygen Concentration

Dimensions:

8.6 in. high x 6.1 in. wide x 3.6 in. deep

(21.8 cm high x 15.5 cm wide x 9.1 cm deep)

Weight:

4.4 lb (2.0 kg); 1.8 lb (.8 kg) optional Airbelt

Universal Power Supply:

AC Power Supply: 100 – 240 VAC (1.0 amp at 120 VAC)
DC Power Supply: 11-16 VDC (6.0 amps max. at 13.5 VDC

nominal)
FreeStyle unit: 28 W nominal at 3 setting


Battery duration:

3 setting - 2 hours; 2 setting - 2.5 hours; 1 setting - 3.5 hours

Optional AirBelt when combined with the internal battery:

3 setting - 5 hours; 2 setting - 6 hours; 1 setting -10 hours

Battery charge time:

3.5 hours; optional AirBelt: 3 hours

Warm-up time:

2 minutes


Battery cycle life:


Approximately 300 cycles, then 80% capacity or below


Audible alarms and

pulse visual indicators:



Start-up – audible and visual (GREEN indicator light); Pulse
flow – visual (GREEN light); Battery condition – battery
level (GREEN indicator lights); Cannula disconnect- audible
and visual (RED alarm light); System overdraw –

audible and visual (RED warning and alarm lights); High and
low pressure – audible and visual (RED alarm light);

Service required: visual (YELLOW alarm light)

Sound:

38 dBA at 1 Setting: 41 dBA at 2 setting: 44 dBA at 3 setting

Temperature range
intended for use:

Operational temperature: 41

o

F to 104

o

F (5

o

C to 40

o

C)

Storage temperature: -4

o

F to 140

o

F (-20

o

C to 60

o

C)