4 freestyle specifications – AirSep MN126-1 C User Manual
Page 10
AirSep Corp. — FreeStyle Service Manual
1-4
1.4 FreeStyle Specifications
Max Operational
Altitude:
up to 12,000 ft (3,657.6 m) (483mmHg) Higher altitudes may affect
performance
* Based on an atmospheric pressure of 14.7 psia (101 kPa) at 70
o
F (21
o
C)
* Operating unit outside of normal operational temperature range can affect performance.
Oxygen Concentration:*
1-3 pulse settings; equivalent to a continuous flow of
90% +5.5/-3% Oxygen Concentration
Dimensions:
8.6 in. high x 6.1 in. wide x 3.6 in. deep
(21.8 cm high x 15.5 cm wide x 9.1 cm deep)
Weight:
4.4 lb (2.0 kg); 1.8 lb (.8 kg) optional Airbelt
Universal Power Supply:
AC Power Supply: 100 – 240 VAC (1.0 amp at 120 VAC)
DC Power Supply: 11-16 VDC (6.0 amps max. at 13.5 VDC
nominal)
FreeStyle unit: 28 W nominal at 3 setting
Battery duration:
3 setting - 2 hours; 2 setting - 2.5 hours; 1 setting - 3.5 hours
Optional AirBelt when combined with the internal battery:
3 setting - 5 hours; 2 setting - 6 hours; 1 setting -10 hours
Battery charge time:
3.5 hours; optional AirBelt: 3 hours
Warm-up time:
2 minutes
Battery cycle life:
Approximately 300 cycles, then 80% capacity or below
Audible alarms and
pulse visual indicators:
Start-up – audible and visual (GREEN indicator light); Pulse
flow – visual (GREEN light); Battery condition – battery
level (GREEN indicator lights); Cannula disconnect- audible
and visual (RED alarm light); System overdraw –
audible and visual (RED warning and alarm lights); High and
low pressure – audible and visual (RED alarm light);
Service required: visual (YELLOW alarm light)
Sound:
38 dBA at 1 Setting: 41 dBA at 2 setting: 44 dBA at 3 setting
Temperature range
intended for use:
Operational temperature: 41
o
F to 104
o
F (5
o
C to 40
o
C)
Storage temperature: -4
o
F to 140
o
F (-20
o
C to 60
o