AirSep MN166-1 B User Manual
Page 26
22
MN166-1
rev
B
How to Respond to FreeStyle 5’s Alarm/Light Indicators
Status
Audible Alarm
Light
Indicates
Action
Indicator
Brief, continuous at
start–up
(Green) pulse;
continuous light
FreeStyle 5 has been
turned on
You may begin to operate your
FreeStyle 5 unit.
Indicator
No
(Green) pulse;
intermittent light
FreeStyle 5 is delivering
oxygen as a pulse flow.
Continue using FreeStyle 5 normally.
Indicator
No
Level indicator
Battery charge level.
Charge as indicated.
Alarm
Continuous:
Beep
(Yellow) alarm;
continuous light
No breath detected by the
unit for a predetermined
time period.
Check the cannula connection. Ensure
that you are breathing through your
nose. If the alarm persists, contact
your Equipment provider.
Alarm
Intermittent:
Beep
25% (Yellow) alarm
BATT; battery indicator
intermittent light
Battery voltage is too low
to operate FreeStyle.
Connect the FreeStyle 5 unit into a DC
outlet or an AC outlet immediately.
Alarm
Intermittent:
Beep, beep
25% (Yellow)
BATT; battery indicator
intermittent light
Battery shutdown due to
low voltage.
Connect the FreeStyle 5 unit into a DC
outlet or an AC outlet immediately.
Alarm
Rapid intermittent:
Beep, beep, beep…
(Yellow) alarm;
intermittent light
Breathing rate is
exceeding the capacity of
the FreeStyle 5 unit.
Reduce activity, then if necessary use
another source of oxygen as available.
Contact your Equipment Provider.
Alarm
Rapid intermittent:
Beep, beep, beep…
(Red) alarm;
continuous light
General malfunction of the
FreeStyle unit has
occurred.
Turn off the unit. Change to another
source of oxygen, and contact your
Equipment Provider.