How to Respond to FreeStyle 5’s Alarm/Light Indicators

Status

Audible Alarm

Light

Indicates

Action

Indicator

Brief, continuous at
start–up

(Green) pulse;
continuous light

FreeStyle 5 has been
turned on

You may begin to operate your
FreeStyle 5 unit.

Indicator


No


(Green) pulse;

intermittent light


FreeStyle 5 is delivering

oxygen as a pulse flow.


Continue using FreeStyle 5 normally.

Indicator

No

Level indicator

Battery charge level.

Charge as indicated.

Alarm

Continuous:

Beep

(Yellow) alarm;

continuous light

No breath detected by the

unit for a predetermined
time period.

Check the cannula connection. Ensure

that you are breathing through your
nose. If the alarm persists, contact

your Equipment provider.

Alarm

Intermittent:

Beep

25% (Yellow) alarm

BATT; battery indicator

intermittent light

Battery voltage is too low

to operate FreeStyle.

Connect the FreeStyle 5 unit into a DC

outlet or an AC outlet immediately.

Alarm

Intermittent:

Beep, beep

25% (Yellow)

BATT; battery indicator
intermittent light

Battery shutdown due to

low voltage.

Connect the FreeStyle 5 unit into a DC

outlet or an AC outlet immediately.

Alarm

Rapid intermittent:
Beep, beep, beep…

(Yellow) alarm;
intermittent light

Breathing rate is
exceeding the capacity of

the FreeStyle 5 unit.

Reduce activity, then if necessary use
another source of oxygen as available.

Contact your Equipment Provider.

Alarm

Rapid intermittent:

Beep, beep, beep…

(Red) alarm;

continuous light

General malfunction of the

FreeStyle unit has

occurred.

Turn off the unit. Change to another

source of oxygen, and contact your

Equipment Provider.