Operating instructions – AirSep MN166-1 B User Manual

Page 21

background image

17

MN166-1

rev

B

[Read the Important Safety Rules section before operating this equipment.]

Operating Instructions

1. Locate and position the FreeStyle 5 so that the air inlets and air outlets are not obstructed.

2. Power the unit from (a) the internal battery; (b) AirBelt; (c) DC outlet (i.e. automobile or motor boat); or (d) an AC

outlet (i.e. normal household electric). (Refer to the Power Supplies section of this Patient Manual.)

This device supplies high-concentration oxygen that promotes rapid burning. Do not allow smoking or open
flames
within the same room of (1) this device, or (2) any oxygen-carrying accessory. Failure to observe
this warning can result in severe fire, property damage, and/or cause physical injury or death.

Do not use your oxygen concentrator in the presence of flammable gases. This can result in rapid burning
causing property damage, bodily injuries or death.

Do not leave a nasal cannula on clothing, bed coverings or chair cushions. If the unit is turned on but not in use,
the oxygen will make the material flammable. Set the I/0 power switch to the 0 (Off) position when the Oxygen
Concentrator is not in use.

Use no oil, grease, or petroleum-based or other flammable products with the oxygen-carrying accessories or the
Oxygen Concentrator. Only water based, oxygen compatible lotions or salves should be used. Oxygen
accelerates the combustion of flammable substances

.

It may be necessary to initially connect the AC or DC power supply to the oxygen concentrator unit before the
unit will operate for the first time on battery power. Your Equipment Provider may have already performed this
step for you.

Charging of the battery may take several minutes after connecting the power to initiate, depending on the
battery’s internal operating temperature. This is a normal condition and is intended for safe battery charging.
This circumstance is more likely when the battery has been fully discharged

.


1. Connect your cannula to the oxygen outlet.
2. Lift the dust cover on the unit.
3. Turn the FreeStyle 5 unit on by pressing the 1, 2, 3, 4 or 5 button on the unit’s keypad for the liter flow prescribed by

your physician. The green LED above the button selected illuminates. Each time you turn on the FreeStyle 5 unit, a brief
alarm sounds. This indicates that FreeStyle 5 is powered for use.

4. To change the pulse flow setting, press the appropriate button. It is normal to hear a difference in sound as you change

the settings.

5. To turn FreeStyle 5 off, press the button corresponding to the pulse flow setting light that is illuminated. When FreeStyle

5 senses inhalation, oxygen is supplied to you through your cannula. The time required to reach minimum oxygen
concentration after turning on the FreeStyle 5 unit is approximately two minutes.