Important safety rules for optional airbelt – AirSep MN166-1 B User Manual

Page 14

background image

10

MN166-1

rev

B

 

 

AirSep offers the OxySafe as an optional accessory. This is intended to be used in conjunction with the FreeStyle 5
concentrator. For customers in regions requiring compliance to EN ISO 8359:1996-Ammendment1:2012, this accessory will
meet this need.

The OxySafe is a thermal fuse to stop the flow of gas in the event that the downstream cannula or oxygen tubing is ignited
and burns to the OxySafe. It is placed in-line with the nasal cannula or oxygen tubing between the patient and the oxygen
outlet of the Visionaire.

For proper use of the OxySafe, always refer to the manufacturer’s instructions (included with each OxySafe kit).

AirSep offers an OxySafe kit that includes OxySafe with 2” of tubing to connect to the oxygen outlet; PN 20629671

Important Safety Rules for Optional AirBelt


 
 

 

For oxygen concentrators equipped with the optional Airbelt :

1. The incorrect use of AirBelt can cause the battery to get hot, ignite, and can cause serious injury. Be sure not to

pierce, strike, step on, or drop the battery, or otherwise subject the battery to strong impacts or shocks.

2. Replace safety cap on AirBelt cord when not in use.

 

 

For oxygen concentrators equipped with batteries:

1. When connected to AC or DC power, the unit’s battery charges until it reaches full capacity, either while the unit is

operating or turned off.

2. If the internal battery fully depletes and the Oxygen Concentrator shuts down, the unit cannot be restarted with the

optional AirBelt. Should this occur, connect your Oxygen Concentrator to its AC or DC power supply for a short
period of time in order to provide sufficient internal battery power to start the unit. AirBelt can then be connected to
provide additional run time.

 

 

For oxygen concentrators equipped with the optional Airbelt:

1. The AirBelt does not need to be fully discharged before recharging. It is recommended to charge it after each use.

2. Depending upon the temperature of the AirBelt, it can take several minutes for the charging cycle to start after

connecting to power. This is a normal condition and is intended for safe charging.