beautypg.com

Transmille 8000 Series User Manual

Page 67

background image

8000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Page 66


The calibration menus will change depending on the selected function :

Function Menus

Available

Notes

DC Voltage
DC Current

Calibration Zero

Calibration +ve Full Scale

Calibration -ve Full Scale

Value

Reset

Set zero calibration point

Set +FS calibration point

Set -FS calibration point

Calibrate FS to set value

Reset range to default points

AC Voltage
AC Current

Calibration 20% FS

Calibration +ve Full Scale

Value

Reset

Set 20% FS calibration point

Set +FS calibration point

Calibrate FS to set value

Reset range to default points


Frequency Response points* :

1kHz (Reference Frequency)**

10Hz
23Hz
40Hz
56Hz

106Hz
206Hz

2kHz

1kHz**

10kHz
20kHz
35kHz
50kHz
75kHz

100kHz
200kHz
400kHz
700kHz

1MHz

* Max. Frequency dependant on range

** 1kHz reference frequency point
MUST be set up first

Resistance 2 Wire
Resistance 4 Wire

Calibration Zero

Calibration +ve Full Scale

Value

Reset

Set zero calibration point

Set +FS calibration point

Calibrate FS to set value

Reset range to default points