Transmille 8000 Series User Manual
Page 50

8000 SERIES OPERATION MANUAL
TRANSMILLE LTD.
Page 49
Thermocouple Measurement : Multimeter Setup (Option)
Press
SHIFT
then DCV to select thermocouple measurement.
Select the thermocouple Type J Type K Type B Type E Type N Type R Type S Type T
using
Г Д buttons or menu dial
Select cold junction compensation mode using
Г Д buttons or menu dial
Auto (PRT Rear) A : Uses the PRT Rear Probe A coefficients - connect to rear panel terminals
Man °C : User measured and entered cold junction temperature value
Additional Settings
Description
DIGITS
Set measurement resolution
FILTER TIME
Set FILTER TIME :
250ms to 64s
Dynamic Mode for Automatic filter setting
based on measurement stability
ANALOGUE FILTER
Set ANALOGUE FILTER mode :
ON / OFF
INPUT TERMINALS
Set INPUT TERMINALS to use :
FRONT / REAR
RATIO
Set RATIO mode :
ON / VALUE / OFF
READING MIN / MAX
Display READING MIN / MAX measurement data
for the selected function / range
TRIGGER
AUTO / MANUAL
GUARD TO LOW
Set GUARD TO LOW mode :
FLOATING / SIGNAL LOW
INTERNAL
TEMPERATURE
Displays the internal temperature