2-Wire Resistance (Electrometer Mode) (Option)
Press
Ω 2WIRE to enter the Ω 2WIRE measurement mode – the status display will indicate the
settings and range information.
Selecting ranges above 10 MOhms will activate electrometer mode - this is displayed in terms of
current (100uA, 10uA, 1uA, 100nA, 10nA). Inputs are measured from the IMeas BNC terminal on the
right hand side of the multimeter. An LED indicates the BNC terminal set is active.
Voltage output on electrometer terminals is configurable from status menu (using menu dial) from
50V to 300V in 50V steps.
Status Display Section
Description
X RANGE
eg. *10m
Ω DC
Display / Selection of 2-Wire Resistance range –
Use menu dial or
RANGE UP RANGE DOWN
buttons to change range.
Selected range indicated by a *
MANUAL
Ranging mode : AUTO or MANUAL
FRONT
FRONT or REAR terminals in use
Additional Settings
Description
DIGITS
Set measurement resolution
FILTER TIME
Set FILTER TIME : 250ms to 64s
Dynamic Mode for filter setting based on measurement stability
ANALOGUE FILTER
Set ANALOGUE FILTER mode :
ON / OFF
READING MIN / MAX
Display READING MIN / MAX measurement
TRIGGER
AUTO / MANUAL
GUARD TO LOW
Set GUARD TO LOW mode : FLOATING / SIGNAL LOW
MATHS, NULL, dB
Set NULL : Store zero offset value
Clear NULL : Clear zero offset value
Set OFFSET : Add a user entered value to the measurement
Set SCALE : Multiply measurement by a user entered value
Clear FORMULA : Clear the formula entered (SCALE & OFFSET)
See dB function for use of Set dB, dB SET ZERO & dBm REF 600 Ohms
POWER / CURRENT
Displays the power dissipation / current flowing
for the specific range & measurement
TEST VOLTAGE
Select Test Voltage 50V • 100V • 150V • 200V • 250V • 300V
for the specific range & measurement
INTERNAL
TEMPERATURE
Displays the internal temperature