8000 SERIES OPERATION MANUAL
TRANSMILLE LTD.
Page 51
PRT Measurement : Multimeter Setup (Option)
Press
SHIFT
then
Ω 4WIRE to select PRT measurement.
Select PROBE CONFIGURATION SINGLE DUAL
Select LINEARISATION IEC 751 ITS-90
Select PRIMARY DISPLAY AVERAGE FRONT REAR
Select PRT Coefficients ITS-90 Front ITS-90 Rear IEC 751 Front IEC 751 Rear
ITS-90 Front
Coefficients
ITS-90 Rear
Coefficients
IEC 751 Front
Coefficients
IEC 751 Rear
Coefficients
R
0
100
R
0
100
R
0
100 R
0
100
a<0°C
0
a<0°C
0
A 3.908299e-3 A 3.908299e-3
b<0°C
0
b<0°C
0
B -5.774999e-7 B -5.774999e-7
a>0°C
0
a>0°C
0
C -4.183e-12 C -4.183e-12
b>0°C 0
b>0°C
0
OK
OK
OK
OK
Reset coefficients
Reset coefficients
Reset coefficients
Reset coefficients
OK
Saves coefficients as entered
Reset Coefficients
Reset coefficient data back to defaults
Additional Settings
Description
DIGITS
Set measurement resolution
FILTER TIME
Set FILTER TIME :
250ms to 64s
Dynamic Mode for Automatic filter (reading stability based)
ANALOGUE FILTER
Set ANALOGUE FILTER mode : ON / OFF
INPUT TERMINALS
Set INPUT TERMINALS to use : FRONT / REAR
RATIO
Set RATIO mode : ON / VALUE / OFF
READING MIN / MAX
Display READING MIN / MAX measurement data
TRIGGER
AUTO / MANUAL
GUARD TO LOW
Set GUARD TO LOW mode : FLOATING / SIGNAL LOW
INTERNAL
TEMPERATURE
Displays the internal temperature