For example : command1/command2 Page 61
Page 61

8000 SERIES OPERATION MANUAL
TRANSMILLE LTD.
Page 60
Programming Commands Overview
The 8000 series is controlled by a set of simple high level commands which can be used either
individually or as part of a command sequence to setup the 8000 Series multimeter to the required
state.
The commands can be joined together using the / (forward slash) character.
The required terminator for the commands to be detected by the multimeter is a carriage return
(ASCII character 13) and should be the last character sent on a command line
For Example :
Command1/Command2
Where each command is represented as Commandx
(x being the command number)
and the carriage return (ASCII character 13) is represented by
All Commands are subject to change or removal.
All commands followed by Carriage Return or Line Feed (ASCII 13 or 10) or both.
Command Response Codes
Commands (excluding Reading commands) response codes
Successful command
*0
Correct command with an incorrect input signal
*2
Correct commands with incorrect parameters
*3
Unauthorised command
*6
Command with incorrect range
*7
Unrecognised command
*9