1 response message for a read request message, 3 response message in the case of an error – TOHO ELECTRONICS TTM000 User Manual
Page 22

22
6.4 Composition of an RTU response message
(transmitted from this product to a high-level computer)
For codes a) through h), see "6.5 Description of RTU codes."
6.4.1 Response message for a read request message
a) Slave address
1BH
b) Function code
03H
d) Number of data items
04H
Number of registers 2
g)
Data for the first register
(a low-level word)
High level
03H
③
When writing ①, ②, ③, and ④
H in the data, write them in the
order described on the left-hand
side. (① represents 1 byte.)
Low level
09H
④
Data for the first register + 1
(a high-level word)
High level
00H
①
Low level
00H
②
e) CRC-16
Low level
B4H
High level
91H
6.4.2 Response message for a write/store request message
a) Slave address
03H
b) Function code
10H
c) Register address
High level
00H
First register address
Low level
00H
d) Number of registers
High level
00H
Fixed at 2
Low level
02H
e) CRC-16
Low level
2AH
High level
40H
6.4.3
Response message in the case of an error
a) Slave address
1BH
b) Function code
83H
In the case of an error, the function
code for the request message + 80H is
entered.
h) Error code
02H
e) CRC-16
Low level
36H
High level
E1H