Page 9: 00 cal

9
SET “OFFSET” (please read before attempting procedure)
Note: The channel must be calibrated before the offset is adjusted.
The ”OFFSET” command allows you to adjust the reading relative to a known
value. To set the ”OFFSET” use the following procedure:
Step 1.
Press and hold the SCAN/adj and HOLD/mode keys for 8 seconds, then
release; the display will alternate between “cal” and “noYes” for ten seconds.
Step 2.
To specify “yes” and select “CH1”, press and hold the HOLD/mode key to
display “CH1” . (To select “CH2”: With “CH1” displayed, press and release
the SCAN/adj key to display “CH2”.)
Step 3.
Press the HOLD/mode key to display “ALAr”.
Step 4.
Press the SCAN/adj key to display “OFFST”.
SCAN
HOLD
adj
mode
noYES
CH1
CH1
SCAN
HOLD
adj
mode
ALAr
SCAN
HOLD
adj
mode
OFF57
ALAr
0.00
CAL
SCAN
HOLD
adj
mode
- Model ASM (8 pages)
- Model 204 (4 pages)
- Model 209 (4 pages)
- Model 227 (8 pages)
- 264 (4 pages)
- 269 (4 pages)
- Model 280E (4 pages)
- Model 370 (48 pages)
- Model 526 (7 pages)
- Model 540 (9 pages)
- Model 595 (8 pages)
- Model 869 (28 pages)
- Model 321 (4 pages)
- Model SRH (12 pages)
- Model 217 (8 pages)
- Model 230 (12 pages)
- Model 242901-06 (1 page)
- Model 265 (4 pages)
- Model 270 (5 pages)
- Model 299 (4 pages)
- Model 470 (35 pages)
- Model 524 (6 pages)
- Model 546 (8 pages)
- Model 730 (16 pages)
- GCT-225 (8 pages)
- Model 141 (4 pages)
- Model 205-2 (4 pages)
- Model 223 (8 pages)
- Model 231 (12 pages)
- Model 256 (8 pages)
- Model 276 (2 pages)
- Model 3100 (8 pages)
- Model 516 (6 pages)
- Model 550 (9 pages)
- Model 760 (18 pages)
- Model LD330 (4 pages)
- Model SRMD (8 pages)
- Model 201 (4 pages)
- Setra Model 206 (4 pages)
- Model 224 (8 pages)
- Model 231RS (16 pages)
- 260 (2 pages)
- 267MR (14 pages)
- Model 278 (2 pages)