Off57 0 – Setra System Model Datum 2000 User Manual
Page 10

10
Step 5.
Press the HOLD/mode key to display the number “0”.
Note: The decimal point that was set during calibration (CAL_U, CAL_C) will
be displayed.
Step 6.
Set the value you wish the meter to display at the presently applied pressure.
Press the SCAN/adj key to move the numbers up (+).
Press the HOLD/mode key to move the numbers down (-).
When the desired number is reached, press the SCAN/adj and HOLD/mode
keys at the same time. The number will flash repeatedly. Then, press the
SCAN/adj key to readjust the number or the HOLD/mode key to accept it.
After the number is accepted, the meter will display “bUSY” for 10 seconds
and then resume normal operation.
SCAN
HOLD
adj
mode
OFF57
0.
SCAN
HOLD
adj
mode
25.
0.
SCAN
HOLD
adj
mode
–25.
0.
See also other documents in the category Setra System Sensors:
- Model ASM (8 pages)
- Model 204 (4 pages)
- Model 209 (4 pages)
- Model 227 (8 pages)
- 264 (4 pages)
- 269 (4 pages)
- Model 280E (4 pages)
- Model 370 (48 pages)
- Model 526 (7 pages)
- Model 540 (9 pages)
- Model 595 (8 pages)
- Model 869 (28 pages)
- Model 321 (4 pages)
- Model SRH (12 pages)
- Model 217 (8 pages)
- Model 230 (12 pages)
- Model 242901-06 (1 page)
- Model 265 (4 pages)
- Model 270 (5 pages)
- Model 299 (4 pages)
- Model 470 (35 pages)
- Model 524 (6 pages)
- Model 546 (8 pages)
- Model 730 (16 pages)
- GCT-225 (8 pages)
- Model 141 (4 pages)
- Model 205-2 (4 pages)
- Model 223 (8 pages)
- Model 231 (12 pages)
- Model 256 (8 pages)
- Model 276 (2 pages)
- Model 3100 (8 pages)
- Model 516 (6 pages)
- Model 550 (9 pages)
- Model 760 (18 pages)
- Model LD330 (4 pages)
- Model SRMD (8 pages)
- Model 201 (4 pages)
- Setra Model 206 (4 pages)
- Model 224 (8 pages)
- Model 231RS (16 pages)
- 260 (2 pages)
- 267MR (14 pages)
- Model 278 (2 pages)