10 Technical data
10.1 Specifications
General
HF62
Device type
Humidity temperature transmitter with analog output signals
Circuit type
Two independent 2-wire, loop powered circuits
Mechanical configuration types
D and W
Power supply and connections
HF62
Supply voltage (VDD)
10…28VDC
V min = 10 V + (0.02 x Load*) *Load resistance in ohms.
Nominal current consumption
(low voltage models)
2 x 20 mA
Cable fittings
Terminal block and M16 cable grip or ½” conduit adapter
Polarity protection
(low voltage models)
Protective diode on V+
Humidity measurement
HF62
Sensor ROTRONIC
Hygromer
®
IN1
Measuring range
0…100 %RH
Measurement accuracy at 23 °C
±1.0 %RH
Repeatability 0.3
%RH
Long term stability
< 1 %RH / year
Sensor time constant
Typical 10 sec, 63% of a 35 to 80 %RH step change (1m/sec air flow at sensor)
Temperature measurement
HF62
Sensor
Pt100 RTD, IEC 751 1/3 class B
Measuring range limits
-100…150 °C / -148…302 °F
Measurement accuracy at 23 °C
±0.2 °C
Repeatability 0.05°C
Long term stability
< 0.1°C / year
Sensor time constant
Typical 4 sec, 63% of a step change (1m/sec air flow at sensor)
Calculated parameters
HF62
Psychrometric calculations
Dew or frost point (user configurable option)