7 operation, 1 internal menu (optional keypad and display), Operation – ROTRONIC HF62 User Manual
Page 16: Internal menu (optional keypad and display), 7operation
E-M-HF62-V1_11
Rotronic AG
Bassersdorf, Switzerland
Document code
Unit
Instruction Manual
Document Type
HygroFlex HF62 Humidity Temperature
Transmitters: User Guide
Document title
Page 16 of 27
© 2008; Rotronic AG E-M-HF62-V1_11
7
Operation
If so desired, use the HW4 software to configure the HF62. Complete the mechanical and electrical installation
and simply power up the HF62.
7.1 Internal menu (optional keypad and display)
IMPORTANT:
The data shown on the LC display of the HF62 as well as the internal menu are generated
exclusively by circuit 1 of the HF62 (humidity measuring circuit).
The temperature data shown on the display as well as settings and functions such as temperature
adjustment, data recording, etc. correspond to the Pt100 RTD used by circuit 1 to compensate the
effect of temperature on the humidity sensor.
Data from circuit 2 (temperature) is not shown on the display and the internal menu of this circuit
is not accessible to the user.
Both the LC display and keypad operate only when circuit 1 (humidity) is powered and the output
loop is closed.
MENU key:
Open / Close menu
ENTER key:
Menu item selection
Menu navigation keys + / - :
Change value up / down
Note: Unauthorized access to the menu can be prevented by disabling the “display menu” setting
(use the HW4 software > Device Manager > Display)