Functions

Factory default

Humidity sensor drift compensation

Disabled

Fail safe mode

Disabled

Simulator mode

Disabled


o

For a detailed description of all AirChip 3000 / HF62 main functions see document E-T-AC3000-DF-V1

o

Instructions regarding the configuration of the HF62 and access to its functions are provided in the
following manuals:

E-M-HW4v2-F2-008
E-M-HW4v2-Main

(§ 6.5)

E-M-HW4v2-DR-001
E-M-HW4v2-A2-001
E-M-AC3000-CP
.

o

The factory default setting for dew / frost point calculation is frost point below freezing

5

Mechanical installation

5.1 General guidelines

Relative humidity is extremely dependent on temperature. Proper measurement of relative humidity requires
that the probe and its sensors be at exactly the temperature of the environment to be measured. Because of
this, the location where you choose to install the probe can have a significant effect on the performance of the
instrument. The following guidelines should guarantee good instrument performance:

a) Select a representative location: install the probe where humidity, temperature and pressure

conditions are representative of the environment to be measured.

b) Provide good air movement at the probe: air velocity of at least 200 ft/ minute (1 meter/second)

facilitates adaptation of the probe to changing temperature.

c) Avoid the following: (1) Close proximity of the probe to a heating element, a cooling coil, a cold or

hot wall, direct exposure to sun rays, etc. (2) Close proximity of the probe to a steam injector,
humidifier, direct exposure to precipitation, etc. (3) Unstable pressure conditions resulting from
excessive air turbulence.

d) Immerse as much of the probe as possible in the environment to be measured.

e) Prevent the accumulation of condensation water at the level of the sensor leads. Install the

probe so that the probe tip is looking downward. If this is not possible, install the probe horizontally.

5.2 HF62 enclosure


The HF62 enclosure consists of a base and a cover held together with 4 screws. To open the enclosure, use a
metric 3 mm hex key. Prior to re-assembling the enclosure, verify that the red seal is sitting properly in its
groove on the base.