5.3 Installation of the enclosure and probe
Mounting position of the enclosure
The enclosure of the HF62 is designed to be mounted in a horizontal position
as illustrated here for the HF62 type W
Mounting hardware for the enclosure (HF62 type W or with cable probe)
Method 1
: The HF62 is supplied with 2 screws, 2 drywall anchors and two
rubber washers. The base of the enclosure has 2 screw-wells (see drawing)
that are normally closed at the bottom. Use the template provided with the
HF62 to drill mounting holes in the wall and insert the drywall anchors. Place
rubber washer on each screw. Insert a screw in each well and push to open the
bottom of th
Method 2
: When a DIN-rail i(35 mm / 1 3/8”) s available use part AC5002 (not
included). This is a DIN-rail mounting kit consisting of 2 clamps that attach to
the back of the enclosure with the screws provid
Mounting hardware for the probe (HF62 type D or with cable probe)
Use parts AC1303-M (compression fitting) and AC1305 (flange) to hold and seal the
probe when mounted through a wall (see Accessories).
o
The enclosure of the HF62 type D does not require any additional support.
o
The enclosure of the HF62 with cable probe can be installed in the same manner as the enclosure of the
HF62 type W.