1-14
Agilent B2200 User’s Guide, Edition 4
Introduction
Specifications
Supplemental Characteristics for B2201A/B2211A
Offset current
1
100 fA, I-V port
IM noise (RMS)
2
5 fA, I-V port
Channel crosstalk capacitance
< 0.5 pF per channel, I-V port
< 3 pF per channel, AUX port
Offset voltage
< 80
μV, I-V port
< 100
μV, AUX port
Settling time
3
2.0 seconds at 300 fA
Bandwidth (at -3dB)
30 MHz, AUX port
Guard capacitance
4
< 145 pF, I-V port
Additional C measurement error
5
< ± 1 % + 0.2 pF, AUX port
1. The offset current when 0 V is applied to all I/O channels. This measurement
is made on a port, 60 seconds after switching a relay.
2. Measured by the Agilent 4156C with the integration time setting 100 PLC
when 0 V is applied to all other paths.
3. The time to settle to within 300 fA of the final value after 10 V is applied.
4. This is guard capacitance for an I/O channel for following conditions. Four
modules are installed in mainframe, and only one I/O channel is connected.
5. For the measurement less than 1000 pF at the frequency 1 kHz to 1 MHz,
using the Agilent 4284A with 3 m cable. After the compensation using the
capacitance compensation algorithm (a function of VXIplug&play driver).