:SYSTem:CTYPe?

For the device reset state, see *RST.

Syntax

:SYSTem:CPON card_number

Example

OUTPUT @Agb2200;":SYST:CPON ALL"

:SYSTem:CTYPe?

This query command returns the ID of the specified card.

Syntax

:SYSTem:CTYPe? card_number

Query response

AGILENT TECHNOLOGIES,model,0,revision <^END>

model = B2210A or B2211A

revision = 1 or later version

If designated card is not installed or initialize failed, this command returns
NONE,NONE,0,0.

Example

OUTPUT @Agb2200;":SYST:CTYPE? 1"

ENTER @Agb2200;A$

:SYSTem:DISPlay:LCD

This command enables/disables the front panel Liquid Crystal Display when the
B2200 is in the GPIB remote mode. The front panel LCD will be automatically set
to ON by the mode transition from GPIB remote to GPIB local.

NOTE

For the high speed switching operation in the GPIB remote mode, disable the LCD.
If the LCD is enabled, the switching speed will slow.

Parameter Explanation

card_number

For Auto Config mode: 0 or ALL
For Normal Config mode: 1, 2, 3, 4, or ALL

Parameter Explanation

card_number

For Auto Config mode: 0
For Normal Config mode: 1, 2, 3, or 4