System:ctype, System:display:lcd – Agilent Technologies Agilent B2201A User Manual
Page 172

5-44
Agilent B2200 User’s Guide, Edition 4
SCPI Command Reference
:SYSTem:CTYPe?
For the device reset state, see *RST.
Syntax
:SYSTem:CPON card_number
Example
OUTPUT @Agb2200;":SYST:CPON ALL"
:SYSTem:CTYPe?
This query command returns the ID of the specified card.
Syntax
:SYSTem:CTYPe? card_number
Query response
AGILENT TECHNOLOGIES,model,0,revision
model = B2210A or B2211A
revision = 1 or later version
If designated card is not installed or initialize failed, this command returns
NONE,NONE,0,0.
Example
OUTPUT @Agb2200;":SYST:CTYPE? 1"
ENTER @Agb2200;A$
:SYSTem:DISPlay:LCD
This command enables/disables the front panel Liquid Crystal Display when the
B2200 is in the GPIB remote mode. The front panel LCD will be automatically set
to ON by the mode transition from GPIB remote to GPIB local.
NOTE
For the high speed switching operation in the GPIB remote mode, disable the LCD.
If the LCD is enabled, the switching speed will slow.
Parameter Explanation
card_number
For Auto Config mode: 0 or ALL
For Normal Config mode: 1, 2, 3, 4, or ALL
Parameter Explanation
card_number
For Auto Config mode: 0
For Normal Config mode: 1, 2, 3, or 4