:DIAGnostic:TEST:CARD:CLEar

:DIAGnostic:TEST:CARD:CLEar

This command clears the relay test result (pass/fail result) of the specified card.

Syntax

:DIAGnostic:TEST:CARD:CLEar card_number

Example

OUTPUT @Agb2200;":DIAG:TEST:CARD:CLE 1"

:DIAGnostic:TEST:CARD[:EXECute]?

This command executes the relay test for specified card, then returns the pass/fail
result. Before starting the relay test, open the input/output terminals (end of cable is
OK, Kelvin cable must be removed). For more information, see “RELAY_TEST”
on page 3-35.

NOTE

This command changes all settings of card to same state as after executing
:SYST:CPON command.

Syntax

:DIAGnostic:TEST:CARD[:EXECute]? card_number

Query response

test_result <^END>

1: failure card exists
0: pass

Example

OUTPUT @Agb2200;":DIAG:TEST:CARD? ALL"

:DIAGnostic:TEST:CARD:STATe?

This command returns the most recent relay test result for the specified card.

Parameter Explanation

card_number

card number: 1, 2, 3, 4, or ALL

Parameter Explanation

card_number

card number: 1, 2, 3, 4, or ALL