Diagnostic:test:card:clear, Diagnostic:test:card[:execute, Diagnostic:test:card:state – Agilent Technologies Agilent B2201A User Manual
Page 149
Agilent B2200 User’s Guide, Edition 4
5-21
SCPI Command Reference
:DIAGnostic:TEST:CARD:CLEar
:DIAGnostic:TEST:CARD:CLEar
This command clears the relay test result (pass/fail result) of the specified card.
Syntax
:DIAGnostic:TEST:CARD:CLEar card_number
Example
OUTPUT @Agb2200;":DIAG:TEST:CARD:CLE 1"
:DIAGnostic:TEST:CARD[:EXECute]?
This command executes the relay test for specified card, then returns the pass/fail
result. Before starting the relay test, open the input/output terminals (end of cable is
OK, Kelvin cable must be removed). For more information, see “RELAY_TEST”
on page 3-35.
NOTE
This command changes all settings of card to same state as after executing
:SYST:CPON command.
Syntax
:DIAGnostic:TEST:CARD[:EXECute]? card_number
Query response
test_result
1: failure card exists
0: pass
Example
OUTPUT @Agb2200;":DIAG:TEST:CARD? ALL"
:DIAGnostic:TEST:CARD:STATe?
This command returns the most recent relay test result for the specified card.
Parameter Explanation
card_number
card number: 1, 2, 3, 4, or ALL
Parameter Explanation
card_number
card number: 1, 2, 3, 4, or ALL