beautypg.com

System:memory:load, System:memory:save – Agilent Technologies Agilent B2201A User Manual

Page 176

background image

5-48

Agilent B2200 User’s Guide, Edition 4

SCPI Command Reference
:SYSTem:MEMOry:LOAD

:SYSTem:MEMOry:LOAD

This command restores the B2200 setup information specified by memory_number.
It must be already saved by “:SYSTem:MEMOry:SAVE”.

Syntax

:SYSTem:MEMOry:LOAD memory_number

Example

OUTPUT @Agb2200;":SYST:MEMO:SAVE 1"

OUTPUT @Agb2200;":SYST:MEMO:LOAD 1"

:SYSTem:MEMOry:SAVE

This command saves the present setup information of the B2200 into the internal
memory specified by memory_number. The previous setup will be deleted. To
memorize the comment for the setup information, use
“:SYSTem:MEMOry:COMMent”.

The following setup information will be saved.

Configuration mode

Connection rule, connection sequence

Bias mode status, bias ports, bias-enabled channels

Couple mode status, couple ports

Ground mode status, ground ports, ground-enabled channels, ground-enabled
ports

Connection status of all switches

Symbol string of all input ports and output ports

Syntax

:SYSTem:MEMOry:SAVE memory_number

Example

OUTPUT @Agb2200;":SYST:MEMO:SAVE 1"

Parameter Explanation

memory_number

memory number: 1 to 8.

Parameter Explanation

memory_number

memory number: 1 to 8.