System:cpon – Agilent Technologies Agilent B2201A User Manual
Page 171

Agilent B2200 User’s Guide, Edition 4
5-43
SCPI Command Reference
:SYSTem:CPON
For B2201A:
•
In the Auto Config mode, "card_description" response is one of following:
• "B2211A 14x12 Low Leakage Switch Module"
• "B2211A 14x24 Low Leakage Switch Module"
• "B2211A 14x36 Low Leakage Switch Module"
• "B2211A 14x48 Low Leakage Switch Module"
•
In the Normal Config mode:
"B2211A 14x12 Low Leakage Switch Module"
Example
OUTPUT @Agb2200;":SYST:CDES? 1"
ENTER @Agb2200;A$
:SYSTem:CPON
This command resets the specified card. Status after this command is shown below:
Channel Configuration:
Not changed.
Connection Rule:
Free
Connection Sequence:
Break Before Make
Bias Mode:
Off
Bias Input Port:
10
Bias-enabled Channels:
All channels (output ports) are bias-enabled.
Couple Mode:
Off
Couple Input Port:
Cleared.
Ground Mode:
Off
Ground Input Port:
12
Ground-enabled Channels:
Cleared.
Ground-enabled Ports:
Cleared.
Card Channel Status:
All relays are opened.
Self-Test Result:
Not changed.
Input Port Symbol String:
Not changed.
Output Port Symbol String:
Cleared.