F-2000 – Blue-White F-2000 – Solvent Weld PVC Tee Fitting User Manual

Page 14

background image

F-2000

Page 14

2.6 Programming The MODEL PC & PB When Setpoints Are Assigned To Batch

Note: The F-2000 must be powered by the AC adapter.

Step 1

Enter the programming mode and assign the setpoint to the BATCH mode.

! Press for at least 1.25 seconds.

! Press to by-pass the first six program screens until you reach the Batch Setpoint Screen (seventh screen).

Step 2

The Setpoint Batch-1 screen, Batch Amount Value is selected.

This screen is used to enter the batch amount value. The relay will energize when the
accumulated total flow equals the programmed value.

Step 3

The Setpoint Batch-2 screen, Auto Batch Start On or Off is enabled.

This screen is used to toggle the auto Batch Start function On and Off.

Note: Select On for auto-Start Batches and Off for manual Start Batches.

Step 4

The Setpoint Batch-3 screen, external equipment Timer amount
(in seconds) is displayed.

This screen is used to enter the number of seconds per batch that the relay will be energized.

Note: Set to zero for manual Start Batch operations.

! Press to select Batch Setpoint ON.

! Press .

! Press to select the digit or the decimal to be modified. The selected digit will blink.

! Press to change the selected digit.

! Press when you have finished.

! Press to select On or Off.

! Press when you have finished.

! Press to select the digit to be modified.

! Press to change the selected digit.

! Press when you’ve finished.

TOTAL

CLEAR

ENTER

CLEAR

SETPOINT

TOTAL

CLEAR

ENTER

TOTAL

CLEAR

ENTER

O

BATCH

SETPOINT

BATCH

SETPOINT

00000

O

BATCH

SETPOINT

2

BATCH

SETPOINT

000

3

CAUTION: A SEPARATE EMERGENCY SHUT-OFF SWITCH FOR EXTERNAL EQUIPMENT IS RECOMMENDED.