Option: Totalizing Mode

16 Series Flow Meters and Controllers can be purchased with the Totalizing Mode option. This option

adds an additional mode screen that displays the total flow (normally in the units of the main flow

screen) that has passed through the meter or controller since the last time the totalizer was cleared.

The Totalizing Mode screen shown below is accessed by pushing the “MODE” button until the label

over it reads “Total”. If your meter or controller is ordered with Totalizing Mode option, pushing the

“Mode” button once will bring up the “Totalizing Mode” display. Pushing “Mode” a second time will bring

up the “Select Menu” display. Pushing it a third time will return you to the Main Mode Sreen.

Hours Mass Clear

0.3 0.00

Mass

SLtr

Air

+0.0 SCCM

Total

Counter – The counter can have as many as six digits. At the time of order, the customer must specify

the resolution of the count. This directly affects the maximum count. For instance, if a resolution of

1/100ths of a liter is specified on a meter which is totalizing in liters, the maximum count would be

9999.99 liters. If the same unit were specified with a 1 liter resolution, the maximum count would be

999999 liters.
Rollover – The customer can also specify at the time of order what the totalizer is to do when the

maximum count is reached. The following options may be specified:
No Rollover – When the counter reaches the maximum count it stops counting until the counter is

cleared.

Rollover – When the counter reaches the maximum count it automatically rolls over to zero and continues

counting until the counter is cleared.

Rollover with Notification – When the counter reaches the maximum count it automatically rolls over to

zero, displays an overflow error, and continues counting until the counter is cleared.
Hours.–.The display will show elapsed time since the last reset in 0.1 hour increments. The maximum

measurable elapsed time is 6553.5 hours (about nine months). The hours count resets when the “clear”

button is pushed, an RS-232 clear is executed or on loss of power
Clear – The counter can be reset to zero at any time by pushing the dynamically labeled “Clear” button

located above the upper right corner of the display. To clear the counter via RS-232, establish serial

communication with the meter or controller as described in the RS-232 section of the manual. To reset

the counter, enter the following commands:

In Streaming Mode: $$T

In Polling (addressable) Mode: Address$$T (e.g. B$$T )