Alicat VC Series Volumetric Flow Controller User Manual
Page 22
22
Once you have established communication with the unit and have a stream of information filling your
screen:
Type *@=A followed by “Enter” (or using the RS-232 communcation select menu, select A as identifier
1.
and exit the screen) to stop the streaming mode of information. Note that the flow of information will
not stop while you are typing and you will not be able to read what you have typed. Also, the unit
does not accept a backspace or delete in the line so it must be typed correctly. If in doubt, simply hit
enter and start again. If the unit does not get exactly what it is expecting, it will ignore it. If the line
has been typed correctly, the data will stop.
You may now poll the unit by typing A followed by “Enter”. This does an instantaneous poll of unit
2.
A and returns the values once. You may type A “Enter” as many times as you like. Alternately you
could resume streaming mode by typing *@=@ followed by “Enter”. Repeat step 1 to remove the
unit from the streaming mode.
To assign the unit a new address, type *@=New Address, e.g. *@=B. Care should be taken not to
3.
assign an address to a unit if more than one unit is on the RS232 line as all of the addresses will be
reassigned. Instead, each should be individually attached to the RS-232 line, given an address, and
taken off. After each unit has been given a unique address, they can all be put back on the same
line and polled individually.
Sending a Set-point via RS-232: To send a set-point via RS-232, “Serial” must be selected under
the “Input” list in the control set up mode. To give controllers a set-point, or change an existing point,
simply type in a number between 0 and 65535 (2% over range), where 64000 denotes full-scale flow
rate, and hit “Enter”. The set-point column and flow rates should change accordingly. If they do not,
try hitting “Enter” a couple of times and repeating your command. The formula for performing a linear
interpolation is as follows:
Value = (Desired Set-point X 64000) / Full Scale Flow Range
For example, if your device is a 100 SLPM full-scale unit and you wish to apply a set-point of 35 SLPM
you would enter the following value:
22400 = (35 SLPM X 64000) / 100 SLPM
If the controller is in polling mode as described in Changing from Streaming Mode to Polling Mode, the
set-point must be preceded by the address of the controller. For example, if your controller has been
given an address of D, the set-point above would be sent by typing:
D22400 followed by “Enter”
To adjust the Proportional and Differential (P&D) terms via RS-232:
Type *@=A followed by “Enter” to stop the streaming mode of information.
To adjust the “P” or proportional term of the PID controller, type *R21 followed by “Enter”.