beautypg.com

Alicat VC Series Volumetric Flow Controller User Manual

Page 30

background image

30

This reduces to: Pa Va / Za Ta = Ps Vs / Zs Ts , eliminating R and n.
Alicat mass flow meters model gas flows based upon the non-ideal gas characteristics of the calibrated

gas. The flow corrections are normally made to 25 C and 14.696 PSIA and the compressibility factor

of the gas under those conditions. This allows the user to multiply the mass flow rate by the density of

the real gas at those standard conditions to get the mass flow rate in grams per minute.
Because we incorporate the compressibility factor into our ‘full gas model’; attempts to manually

compute mass flows from only the P, V, and T values shown on the display will sometimes result in

modest errors.

Note: Although the correct units for mass are expressed in grams, kilograms, etc. it has become standard

that mass flow rate is specified in SLPM (standard liters / minute), SCCM (standard cubic centimeters

/ minute) or SmL/M (standard milliliters / minute).

This means that mass flow rate is calculated by normalizing the volumetric flow rate to some standard

temperature and pressure (STP). By knowing the density at that STP, one can determine the mass flow

rate in grams per minute, kilograms per hour, etc.

STP is usually specified as the sea level conditions; however, no single standard exists for this

convention. Examples of common reference conditions include:

0°C and 14.696 PSIA

25°C and 14.696 PSIA

0°C and 760 torr (mmHG)

70°F and 14.696 PSIA

68°F and 29.92 inHG

20°C and 760 torr (mmHG)

MC Series Flow Controllers reference 25ºC and14.696 PSIA (101.32kPa) - unless ordered

otherwise and specified in the notes field of the calibration sheet.

Standard Gas Data Tables: Those of you who have older Alicat products (manufactured before

October 2005) may notice small discrepancies between the gas property tables of your old and new

units. Alicat Scientific, Inc. has recently incorporated the latest data sets from NIST (including their

REFPROP 7 data) in our products’ built-in gas property models. Be aware that calibrators that you may

be spot checking against may be using older data sets such as the widely distributed Air Liquide data.

This may generate apparent calibration discrepancies of up to 0.6% of reading on well behaved gases

and as much as 3% of reading on some gases such as propane and butane, unless the standard was

directly calibrated on the gas in question. As the older standards are phased out of the industry, this

difference in readings will cease to be a problem. If you see a difference between the Alicat meter and

your in-house standard, in addition to calling Alicat Scientific at (520) 290-6060, call the manufacturer

of your standard for clarification as to which data set they used in their calibration. This comparison will

in all likelihood resolve the problem.