Instruction manual

Alpha DO 2000

− If the proportional controller type is set to pulse frequency control (PFC):

9. Selecting the relay status under Non-Alarm condition (REL): Press

▲ or ▼ key to choose the desired relay status (de-energized =
“DEEN” or energized = “EN”). Press ENT key to confirm your selection.

10. Setting the proportional range (PRP): Press

▲ or ▼ key to set the

proportional range (setting range: 10 to 500%). Press ENT key to
confirm your setting.

11. Setting the pulse frequency (F.PF): Press

▲ or ▼ key to set the

pulse frequency (setting range: 60 to 120 pulses per minute). Press
ENT key
to confirm your setting.

12. Setting the integral action time (IAT): Press

▲ or ▼ key to set the

integral action time (setting range: 0.0 to 999.9 minutes). Press ENT
key
to confirm your setting.

The controller goes back to “CNTR” screen.

Continue with rest of the Setup procedures by pressing

▲ or ▼ key or return

to the Measurement mode by pressing

▲ and ▼ keys (escape)

simultaneously.