Setup mode, 1 enter setup mode, Enter setup mode – Thermo Fisher Scientific Alpha DO 2000 User Manual
Page 27: Ofs 13.0, Instruction manual, Alpha do 2000

Instruction manual
Alpha DO 2000
6. Setup Mode
6.1 Enter Setup Mode
In the Setup mode the transmitter can be configured to your individual requirements.
MEAS
SETUP
SETUP
SETUP
SETUP
SETUP
SE TUP
SETUP
SETUP
SETUP
HOLD
HOLD
HOLD
HOLD
HOLD
HOLD
HOLD
HOLD
HOLD
%
°C
Temperature settings
see section 6.3
Relay A (set point 1) settings
see section 6.4
Relay B (set point 2) settings
see section 6.4
Controller settings
see section 6.5
Current output 1 settings
see section 6.6
Current output 2 settings
see section 6.7
Wash relay settings
see section 6.8
Calibration
see section 6.10
SETUP
HOLD
Configuration settings
see section 6.9
OFS
13.0
DO
Offset settings
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
see section 6.2
CAL
1. While in Measurement mode press ENT key.
2. The display prompts you to enter the security code.
Set the security code with
▲ or ▼ key to:
− “SCD 22” if you want to change parameter
settings
− “SCD 000” if you only want to view parameter
settings (view only mode)
3. Press ENT key.
Î NOTE: To exit Setup mode at any time press ▲ and ▼ keys simultaneously
(escape). The transmitter returns to the Measurement mode.
19