Instruction manual

Alpha DO 2000

1. From the setup mode, select the “CNTR” screen and then press ENT key.

2. Selecting the controller type (TYP): Press

▲ or ▼ key to select the

suitable controller type:

− OFF = controller off

Use control OFF to operate controller as a monitor only or to prevent
relays from switching. When the controller is set to OFF, all the LEDs
will be switch off in the Measurement mode

− L.CT = limit value control (on/off control).

Use limit control with pumps or valves for fast response

− P/PI = proportional/integral control

Use proportional control to operate your pumps smoothly or for precise
control of proportioning valves. Use PI controller to eliminate steady
state error.

Î NOTE: Please refer to Appendix 3 for detailed information on controller

settings.


Press ENT key
to confirm your selection.

− If the controller is set to limit value control (L.CT):

3. Selecting the relay status under Non-Alarm condition (REL): Press

or

▼ key to choose the desired relay status (de-energized = “DEEN” or

energized = “EN”). Press ENT key to confirm your selection.

− If the controller is set to proportional control (P/PI):

4. Selecting the proportional controller type (TYP): Press

▲ or ▼ key to

select the suitable controller type (“PLC” = pulse length control, “PFC” =
pulse frequency control). Press ENT key to confirm your selection.

− If the proportional controller type is set to pulse length control (PLC):

5. Selecting the relay status under Non-Alarm condition (REL): Press

▲ or ▼ key to choose the desired relay status (de-energized = “DEEN”
or energized = “EN”). Press ENT key to confirm your selection.

6. Setting the proportional range (PRP): Press

▲ or ▼ key to set the

proportional range (setting range: 10 to 500%). Press ENT key to
confirm your setting.

7. Setting the pulse length (T.PL): Press

▲ or ▼ key to set the pulse

length (setting range: 0.5 to 20 seconds). Press ENT key to confirm
your setting.

8. Setting the integral action time (IAT): Press

▲ or ▼ key to set the

integral action time (setting range: 0.0 to 999.9 minutes). Press ENT
key
to confirm your setting.