Alpha DO 2000

3.2 Connection Diagram

Wall mount version


Connections Description

1.

AC mains live wire

14.

Hold function

2.

AC mains neutral wire

19.

12V output power supply

3.

AC mains protective earth wire

20.

12V output ground

4.

Alarm relay (NO)

21.

Earth ground

5.

Alarm relay common

22.

DO input, Anode

6.

Alarm relay (NC)

23.

DO input, Cathode

7.

Wash relay

24.

Temperature – T2

8.

Wash relay

25.

Temperature – N.C

9.

Relay B (SP 2)

26.

Temperature – T1

10.

Relay B (SP 2)

28.

4-20 mA temperature output, -ve

11.

Relay A (SP 1)

29.

4-20 mA temperature output, +ve

12.

Relay A (SP 1)

30.

4-20 mA DO output, -ve

13.

Hold function

31.

4-20 mA DO output, +ve

Probe Cable Connections

Cable Connection

Controller/Transmitter
Terminal

Colorless/Center (Cathode)

K (23)

Brown (Anode)

A (22)

Blue (T1)

TEMP (26)

Yellow (T2)

TEMP (24)

Green/Yellow (Cable Outer Shield)

A (22)

Connect jumper wire

From E(21) to A (22)