Getting started, 1 description of instrument, Description of instrument – Thermo Fisher Scientific Alpha DO 2000 User Manual
Page 11

Instruction manual
Alpha DO 2000
2. Getting Started
2.1 Description of Instrument
The Alpha DO 2000 is used for measuring dissolved oxygen (in % saturation,
mg/l or ppm) and temperature values. The dissolved oxygen values can be
measured using limit or P/PI control.
The transmitter is available in two versions, one for panel mounting and one
for wall mounting in a enclosure. The transmitter can be used for applications
such as water treatment and monitoring, chemical and food processing,
clean and wastewater control and fish farming.
This controller/transmitter has many user-friendly and safety features which
include:
• Menu-driven program that simplifies set-up
• Built-in non-volatile memory to ensure that calibration and other
information are not erased if power supply fails.
• Push-button for calibration and sensor offset adjustment from the keypad.
• Automatic
temperature
compensation (ATC).
• Manual temperature compensation setting without the ATC probe, with
independent setting for calibration and process temperature.
• Two galvanically isolated current outputs 0/4...20mA.
• 0 to 1999 seconds time delay adjustment on all relays – minimizes false
alarms.
• Separately adjustable high and low set-point hysteresis (dead bands)
prevent chattering of relays around the set points.
• Three control modes: limit controller, P controller and PI controller (P/PI
controller as pulse length or pulse frequency).
• Large dual display LCD for easy reading with clear multiple annunciators,
alarm status, operational and error messages.
• Two switching contacts as set-point relays.
• Separate alarm relay alerting you to set point limits exceeded for a certain
time and if the NTC wires are broken or disconnected during the ATC
function.
• Wash relay.
• Hold function to freeze output current (0/4...20mA) and release control
relays.
• LED indicators signal control activities to visually monitor transmitter
status from a distance.
• Protection against electromagnetic interference.
• Back lit and UV light protected LC display.
3