Instruction manual

Alpha DO 2000

5.3 View Probe Condition

The probe condition is an indication of its current working condition with
respect to its original condition. The probe condition is calculated and stored
in the controller each time calibration is done.

You can view probe condition without entering into calibration mode.

89.5

24.5

ADJ

C.CD

100.0

000

ATC

MEAS

%

0.0

CAL

ENT

ENT

ENT

CAL

1

2

4

6

95.5

CAL

5

ENT

CAL

%

DO

CAL

3

CAL

ENT

1. While in Measurement mode press CAL key.

2. The display prompts you to enter the calibration security code (CCD). Leave

the security code at “000” (view only mode). Press ENT key.

3. The display shows “CAL DO”. Press ENT key.

4. The display shows the zero point current in nA. Press ENT key.

5. The display shows the current in nA for the second calibration point. Press

ENT key.

6. The display shows the selected measuring range in %. Press ENT key to

return to the Measurement mode.

Î NOTE: If no key is pressed for 30 seconds during the above steps, the

controller automatically returns to the Measurement mode.

18