Calibration (CAL) sub-function

The calibration procedure in Advanced Set-up mode is identical to the procedure in
the Calibration mode. The only difference is that the controller will revert back to
Set-up mode (instead of Measurement mode) after calibration is completed.


7.9.1

Entering Calibration mode from Advanced Set-up mode

1. Enter Advanced Set-up mode. Push the ENTER key and scroll to

Advanced Set-up security code “22”. Push the ENTER key again.

2. Press

the

or keys to scroll until the upper display shows “CAL”.

3. See section 6 for complete calibration procedures.