7.7.4

Selecting Dissolved Oxygen value at 20mA

This parameter lets you choose the Dissolved Oxygen value at which the
transmitter output will be 20mA.

1. Follow directions in 7.7.1 to enter Current Output mode.

2. Press

the ENTER key until the upper display shows a Dissolved Oxygen

value and the lower display shows “r.20”.

3. Press

the

or keys to select dissolved oxygen value to be equivalent

to 20 mA (Default is 8.00mg/l or 100%).

4. Press

the ENTER key to confirm your selection and return to Advanced

Set-up mode, or return to Measurement mode by pressing the

and

keys simultaneously (escape).