7.7.4
Selecting Dissolved Oxygen value at 20mA
This parameter lets you choose the Dissolved Oxygen value at which the
transmitter output will be 20mA.
1. Follow directions in 7.7.1 to enter Current Output mode.
2. Press
the ENTER key until the upper display shows a Dissolved Oxygen
value and the lower display shows “r.20”.
3. Press
the
▲ or ▼ keys to select dissolved oxygen value to be equivalent
to 20 mA (Default is 8.00mg/l or 100%).
4. Press
the ENTER key to confirm your selection and return to Advanced
Set-up mode, or return to Measurement mode by pressing the
▲ and ▼
keys simultaneously (escape).
7.8
Configuration (ConF) sub-function
This group of parameters lets you configure the controller to suit your requirements.