760 pa – Thermo Fisher Scientific Alpha DO 1000 User Manual

Page 24

background image

16

HOLD

SETUP

760

pa

HOLD

SETUP

HOLD

SETUP

8. Press

the

▲ or ▼ keys until the Dissolved Oxygen value coincides with

the correct Dissolved Oxygen value noted in Step 2. As the

▲ or ▼ keys is

pressed to adjust the Dissolved Oxygen value, you will see that the ‘offset’
value in its upper display also changes.

9. Press the ENTER key to accept the offset value. Note: The offset value is

reset during full calibration (See section 6).

10. Proceed to input Salinity values or return to Measurement mode by

pressing the

▲ and ▼ keys (escape) simultaneously.

7.2

Setting the Salinity Value

1. Enter Advanced set-up mode. Push the ENTER

key and scroll to Advanced Set-up security
code “22”. Push the ENTER key again.

2. Press

the

▲ or ▼ keys to scroll until the upper

display shows OFS.

3. Press the ENTER key till the upper display

shows the “salinity value” in ppt, while the lower
display shows “SAL”. (This function is not available in the % mode and
this window is not displayed
)

4. Press

the

▲ or ▼ keys until the salinity value (between 0.0 ppt and 50.0

ppt) coincides with the correct salinity value.

5. Press the ENTER key to accept the Salinity value. The controller

automatically corrects for the final Dissolved Oxygen measurement.

6. Proceed to Section 7.3 to input the Pressure values or return to

Measurement mode by pressing the

▲ or ▼ keys (escape) simultaneously.

7.3

Setting the Pressure Value (mmHg or Pascal)

1. Enter Advanced set-up

mode. Push the ENTER
key and scroll to
Advanced Set-up
security code “22”.
Push the ENTER key
again.

2. Press

the

▲ or ▼ keys to scroll

until the upper display shows OFS.

3. Press the ENTER key till the upper display shows

“Hg” and the lower display shows “bar”.

4. Press

the

▲ or ▼ keys to select the units of

pressure: if in mmHg, upper display shows “Hg”
and lower display shows “bar”; or in Pascal, upper
display shows “Pa” and lower display shows “bar”.

5. Press the ENTER key to accept the units of measure for pressure. Use the

▲ or ▼ keys to scroll the correct pressure.

HOLD

SETUP

ppt

0.0

sal