Thermo Fisher Scientific CyberScan pH 1500 User Manual

Page 44

background image

Instruction Manual

CyberScan pH1500

42

6.3 Temperature Calibration

1.

Press mode to go to the absolute mV measure screen

2.

Press std. this will take you to the temperature calibration (MTC or ATC)

3.

Press p or q to adjust the temperature.

4.

Press enter to confirm the reading.


Note: Maximum adjustable temperature offset is ±5°C/ 9°F.