beautypg.com

Thermo Fisher Scientific CyberScan pH 1500 User Manual

Page 16

background image

Instruction Manual

CyberScan pH1500

14


4.1.2 P2.0 View pH Electrode Slope

This setup menu allows you to view the pH electrode slope in % after
2-point standardization is successfully done.

pH

View

press

to set value

press

to

accept

enter

ELECTRODE

% SLOPE


To View pH Electrode Slope

1. Access the View pH Electrode Slope menu from the pH Measure

screen by pressing the setup key.

2. Press

▲ and ▼ keys to scroll until display shown as above.

save

view

3. Press

enter to view the offset. A series of dashes “- - -“ will be

displayed if the unit has not been standardized

pH

% SLOPE