4.1.3 P3.0 pH Buffer Group

This setup option allows you to select from 3 different buffer groups,
each containing 5 buffers, for auto buffer recognition.

The 3 buffer groups are:

USA buffers:

2, 4, 7, 10 and 12

European buffers:

1, 3, 6, 8 and 10

NIST buffers:

1.68, 4.01, 6.86, 9.18 and 12.46




To Set pH Buffer Group

1. Access the Select Buffer menu from the pH Setup Screen by

pressing the setup key.

press

to set value

press

to

accept

enter

ELECTRODE

BUFFER
select


2. Press

▲ and ▼ keys to scroll until display shown as above.

save

view

.