rating plate (located on the back of the unit). Ensure that the voltage selector switch (located above
the On/Off switch) is set to the appropriate voltage for the local supply. Please note that the unit
must be earthed to ensure proper electrical safety. Connect the mains cable to a suitable plug
according to the table below.
Connections
220V-240V, 50/60Hz Supply
110V-120V, 50/60Hz Supply
Live
brown
black
Neutral
blue
white
Earth
green/yellow
green
3. UK ONLY: The fused plug supplied with the mains cable is fitted with a 10 Amp fuse to protect the
instrument and the user.
4. Units showing 230V, 50/60Hz on the rating plate also operate between 210 and 260V; units with
120V, 50/60Hz also operate between 100 and 130V and units marked with 100V will operate
between 90 and 110V. In all cases the heating rate will degrade by approximately 8% at the
extremes of the voltage range.
5. Plug the mains cable into the socket on the back of the instrument.
6. Place the unit on a suitable flat bench or in a fume cupboard if required, ensuring that the air inlet
vents on the underside are free from obstruction. Position the instrument with a minimum distance
all round of 200mm from walls or other items and between each unit if multiple units are being
used to permit efficient air flow for each instrument.
7. Switch on the instrument:
I
mains switch On
O mains switch Off
REPLACEMENT CABLE
Should the mains lead need replacement, a cable of 1mm
2
of harmonized code H05VV-F connected to
an IEC320 plug should be used. IF IN DOUBT CONSULT A QUALIFIED ELECTRICIAN.
WORKING CONDITIONS
The thermal cycler is designed to operate under the following conditions:
• Indoor use
• Ambient temperature range +5°C to +40°C
• Altitude to 2000m
• Mains supply fluctuations not exceeding 10%
• Over voltage category II IEC 60364-4-443
• Pollution degree 2
Note: The control specifications are quoted at an ambient temperature of 20°C. The specification may
deteriorate outside an ambient temperature of 10°C to 30°C.