beautypg.com

Techne TC-512 User Manual

Page 36

background image

34

OPTION 3: CREATING A NEW PROGRAM

The final option is to create a completely new
program by selecting "New" in the Select
Program menu. This section also contains
details of the individual parameters and
functions.

To create a new program, select "New" in the
Select Program menu.

Type in a program name; for details see the
previous page.

The cursor icons can be used to scroll up and
down to access more parameters. Scrolling up
the menu displays the default parameters:

• Heated lid
• Pause before the program
• Initial denaturation
• Hot start
• Final extension
• Final hold

The default parameters are defined in the
System user menu and information about the
programming of these parameters can be
found on page 41. All of the default
parameters can be set uniquely within an
individual program.

The keys in the editing menu are as follows:

"Del"

Delete will remove only the highlighted step. The default parameters cannot be deleted.

"Ins"

Insert will add a next step if no "Next step" is present.

"Save As"

Allows the program to be saved with an alternative name.

"Save"

Saves the program.

"PWord"

Adds a password which protects the program so other users cannot change it.

"Exit"

This returns the screen to the Select Program menu.

Select Program

14526 bytes of memory free

b 2 STEP PCR TEMPLATE

b 3 STEP PCR TEMPLATE

b ICE BUCKET

b LITIGATION 15 DEGREES

b RT PCR TEMPLATE

New

Exit

▼ ▼

Editing

B ACTIN

Initial denaturation

Disabled

Hot Start

Disabled

Next Step

Next Step

Final extension

Disabled

Final hold

Disabled

Exit

Save

Save As

PWord

Del

Ins

▼ ▼

The Next step line is inserted automatically
between the Hot Start and Final extension.
Select "Next Step" to open the Edit program
function menu. If Next Step is not displayed
then selecting "Ins" will add a Next Step
below the highlighted line.

TC512 Techne Manual 19/12/08 3:30 PM Page 36