49
To format a new SmartMedia card;
1. Open and lift the heated lid of the thermal
cycler and then lift the silver lid on the front
of the unit.
2. Remove any memory card already present in
the slot by pressing the button on the left-
hand side of the slot.
3. Then insert the new memory card as
described on page 24.
4. Once the card is inserted correctly then
press "System Card" on the Card Utilities
screen.
The screen will then display an "Erasing block
xxxx" message.
The erasing block message will then change to:
Remove the formatted card and press "Exit".
SYSTEM CONFIGURATION
This contains all the settings for programming the instrument such as units of temperature, date
format, language, and date and time settings. Other important features are enabling the auto restart
in case of a power failure and setting the comms address of the instrument if multiple units are linked
to a computer. This part of the software also contains information relating to the cycle count for the
block and the heated lid.
The cursor buttons can be used to scroll up
and down the menu to access additional
parameters.
A parameter is selected by touching the box
on the screen. Then, depending on the
parameter the screen will either toggle
between the options or display the numerical
interface over the page.
Bibby Scientific
TC-512 TEST UNIT
Programming SmartMedia
Erasing block xxx
System Card
Erase
Card
Exit
Programming SmartMedia
Programming xxxx
SmartMedia successfully programmed
Remove programmed card
System Configuration
Temperature units
°C
Beeper
On
Comms Address
01
Language
English
Date format
dd/mm/yyyy
Exit
▼ ▼