Techne TC-512 User Manual
Page 22

TC-512 SPECIFICATION
Temperature
Gradient Range
20°C to 70°C
Maximum Gradient
30°C
Minimum Gradient
1°C
Number of temperature probes
4
Block temperature range
4°C to 99°C
Typical heating rate
3.0°C/sec
Typical cooling rate
1.3°C/sec
Block uniformity (without gradient)
< +
_ 0.3°C
Column uniformity with a 10°C gradient
< +
_ 0.3°C
Control accuracy
< +
_ 0.3°C
Gradient accuracy
< +
_ 0.3°C
Temperature set point adjustment
+
_ 0.1°C
Heated Lid
The heated lid is only operational if the block temperature is set above 35°C
Selectable heated lid temperature
100°C to 115°C
Heated lid enable/disable
Yes
Warm up time from ambient
4 minutes
Over-temperature cut-out
Yes
Regulated heated lid pressure
Yes
Programs
Password protected
Yes
Incremental/decremental hold time
Yes
Incremental/decremental temperature
Yes
Variable programmable ramp rate steps
+
_ 0.1°C
Programmable ramp rate
0.1°C
Gradient calculator
Yes
Maximum programs stored internally
84
Maximum number of segments per loop
>60
Maximum number of loops per program
30
Maximum programs stored externally
500
Program naming capacity
Alpha numeric
Program interface
Touch screen
Pause facility
Yes
End of program alarm
Yes (can be disabled)
Auto restart on power failure
Yes
Run completion time
Yes
20
Technical Specification at an ambient temperature of 20°C
TC512 Techne Manual 19/12/08 3:30 PM Page 22