Techne TC-512 User Manual
Page 46

44
The data log contains information relating to the events which occurred during the program. The
recorded events and descriptions are shown below:
Start: Start of program event is generated when the run button is first pressed.
Stop: Created by pressing the "Stop" button to stop a run prematurely.
Start of program: Created whenever a new program starts.
Start of cycle: Created when the current program starts a new cycle.
Start of segment: Created when a new segment (step) starts.
Start of hold time: Created when the segment reaches the hold window.
End of program: Created when the program finishes.
Pause on: Created when the "Pause" button is first pressed to pause a program.
Pause off: Created when the "Unpause" button is subsequently pressed to end a pause.
TC-512 program
30/11/98
14:27:25
System Configuration
System Version:
0.49 1.00
Unit Serial No:
88192/01
Last System Test:
19/11/1988 09:34:24
Program:
DEPT PROGRAM
Date last edited:
12/10/1988 17:46:11
Data log file:
B ACTIN
Event
Date & Time
Func
Cycle
Step
Temp
Set point
Ramp Rate
Program
30/11/98 14:27:26
01
01
Ramping
30/11/98 14:27:26
01
01
01
4.0°C
4.0°C
- MAX °C/s
Holding
30/11/98 14:27:51
02
01
01
6.3°C
4°C
Ramping
30/11/98 14:37:52
02
01
01
4.0°C
95.0°C
+ MAX °C/s
Holding
30/11/98 14:38:29
02
01
01
88.1°C
95.0°C
Ramping
30/11/98 14:38:59
02
01
02
95.0°C
55.0°C
- MAX °C/s
Holding
30/11/98 14:39:26
02
01
02
59.2°C
55.0°C
Ramping
30/11/98 14:39:56
02
01
03
55.0°C
72.0°C
+ MAX °C/s
Holding
30/11/98 14:40:05
02
01
03
65.5°C
72.0°C
Ramping
30/11/98 14:41:05
02
02
01
72.0°C
95.0°C
+ MAX °C/s
Holding
30/11/98 14:41:16
02
02
01
88.5°C
95.0°C
Signed:
Example of the Data Log
Data Log:
B ACTIN
Program Name:
DEPT PROGRAM
Time Started:
30/11/1998 14:27:26
Temperature (°C)
10
20
30
40
50
60
70
80
90
15
10
5
Typical graph