Metex Shark Multi-Parameter Controllers & Analyzers User Manual

Page 4

background image

M

/cm

3

0 to 19.99 0.01

uS/cm

3

0 to 2.000 0.01

0 to 20.00 0.1

0 to 200.0 0.1

0 to 2000

1.0

mS/cm

3

0 to 20.00 10

0 to 200.0 50

Section I - Specifications

Page 2

S

H

A

R

K

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

Section I - Specifications

pH

ORP

Conductivity

Flow

Display

Front Panel: 4 x 7 segment 1/2” LED display, 1 LED indicator 0n-line,7 LED Bar Graph
Inside Panel: 2 x 16 alpha-numeric LCD display

Power Requirements

120Vac (±10%) 50/60Hz (less than 12VA) or 240Vac (±10%) 50/60Hz (less than 12VA)

Flow:

0 to 9999 with selectable
flow rate units

Volume: 0 to 9999 with Auto Range

Measuring Range

pH:

0.01 to 14.00

ORP: -1999 to +1999mV
(Dependent on sensor)

Flow rate units: Gallons (GP), Cubic

Temp: 0 to 100°C or

Temp: 0 to 100°C or

Feet (CF), Liters (LP), Cubic Meters

32° to +212°F

32° to +212°F

(CM), custom by entering factor
related to Gallons

Time units: Seconds (S),

Temp:

0 to 100°C or 32° to +212°F

Minutes (M), Hours (H)

Automatic or Manual

Temperature

Automatic or Manual

Not required

User selectable temperature

Not required

Compensation

0 to 100°C (32° to +212°F)

compensation slope 0.0 to 10.0%/°C.
0 to 100°C (32° to +212°F)

Temperature Unit

°C or °F

Not required

Temperature Sensor

User selectable: 300

NTC Thermistor, 3000

NTC Thermistor or Pt. 1000 RTD

Not required

Auto-Calibration

Manual Calibration

Dry Calibration

Calibration Modes

Manual Calibration

Temperature Calibration

Sample Calibration

K factor Input

Temperature Calibration

Temperature Calibration

Ambient Conditions

Temperature: -20°C to +60°C or -4°F to +140°F Humidity: 0 to 90% RH (non-condensing)

Menu Access

Auto-Calibration, Manual

Manual-Calibration,

Manual Calibration

Front Panel

Calibration,Temperature

Temperature Display

Temperature Display

Not available

Display

Menu Access Inside Panel

Full Access to all parameters of operations menu

Sensor to SHARK

Differential Sensor: 3000 ft

300 ft

2000 ft

Distance

Combination Sensor: 10 ft
Two Control Relays, 10A / NO, 5A / NC @ 240VAC or 28VDC.
Mode: Process control, Adjustable parameters: process direction,(rising or falling) on-set-point, off set-point,
(0 to 100% of full scale), cycle timer (on / off, 0 to 600 seconds), failsafe (on / off).

Relay Outputs

One Alarm Relay, 10A / NO, 5A / NC @ 240VAC or 28VDC.
Mode: High / Low Alarm, Adjustable parameters: Low on / Low off set-point (0 to 100% of full scale, low on must be less than
low off), High On / High Off set-point (0 to 100% of full scale, High on must be greater than High off).
4 to 20mA Channel 1
Isolated Output, Range expand 0 - 100% of full scale (min segment 10% of full scale), max. load 800

Analog Outputs

4 to 20mA Channel 2
Isolated Output, Range expand 0 - 100% of full scale (min segment 10% of full scale), max. load 800

Can be set to track temperature if sensor is equipped with a temperature sensor

Memory Back-up

All user settings are retained indefinitely in memory (EEPROM)

Mechanical

Enclosure: NEMA 4X, 1/4 DIN, polycarbonate enclosure with four 1/2” conduit holes
Mounting: Universal Mounting kit for surface, pipe and panel mount, is included

Sensor Input

Probe: -600 to +600mV

Probe: -1999 to +1999mV

Cell: 0 to 9999

Paddle: 0 to 2000Hz

Temp. Sensor: 0 to 9999

Temp. Sensor: 0 to 9999

Temp. Sensor: 0 to 9999

Invalid Entries

Invalid entries cannot be stored

Manual Test Mode

Process value can be simulated with arrow keys to verify correct setup of outputs

Manual Relay Override

Relays can be set to on / off / auto, to verify correct wiring of auxiliary devices, or to manually adjust process

Output Hold

All outputs are placed on hold when SHARK is in Menu mode

Recall data from last

Recall data from last calibration, calibration mode,

calibration, calibration buffer

Calibration Data

1st & 2nd accepted buffer value and probe mV output,

accepted value, and cell

Recall store K factor.

calibration temperature, calibration slope, and probe

resistance, calibration

efficiency

temperature

Auto Return

User selectable auto return if SHARK is left in menu mode or if relays are left in manual override mode for more than 10 min.

Display Damping

User can select rate at which SHARK updates display. Enables display damping of unstable process

Net Weight

2.2lbs (1kg)

Approvals

ULC (pending)