6) Confirm erasure of all logged data and set to Demand
mode
• Main display: CLEAr
• Secondary display: CNFRM
• the ▼ and ▲ keys are ignored.
• press the ENTER key to erase the data, set Demand mode, and
exit configuration after briefly displaying CLr to confirm erasure.
• press the CONFIG key to exit configuration without erasing the
data.
While not logging data, the gauge operates as described in the user
manual received with the gauge.
When a log run is started, the pressure unit and custom factor are left
at the values set by the existing configuration functions.
While logging data, the following changes occur:
• Pressing the power button causes the log run to be stopped
before power is turned off.
• Forced power off on low battery detection causes the log run to
be stopped before power is turned off.
• The zero key is active and an event record (including the change
in the zero offset value) is recorded in the log.
• Auto off is disabled. It is restored to its original value at the end
of the run.
• The engineering units, damp, sample rate, tare, and custom
factor items on the existing configuration menu are disabled.
• The units field on the display cycles between the units, the text
LOG, and the text RUN at a rate of once per second. This
provides visual feedback that data logging is active, and the
number after the text RUN identifies the run number for later
reference.
4.
BetaLOG Microsoft Windows Application
This application provides the following features:
• set up data logging configuration in a gauge
• upload logged data from the gauge, including export of run data to
text, CSV, and Excel format files
• clear logged data from the gauge
The application requires Microsoft Windows 2000 (SP4), XP (SP2), or
Vista. Exporting data to Excel files requires Microsoft Excel version
2002 or later.