beautypg.com

Martel Electronics BetaLOG-PI User Manual

Page 10

background image

2) Display number of free log records

• Main display: number of free records

• Secondary display: FREE

• the ▼ and ▲ keys are ignored.

• press the CONFIG key to step to the next menu item if not

logging, otherwise it exits configuration.

• press the ENTER key to exit configuration.

3) Select logging interval

• Main display: interval choice presently selected, one of 1, 2, 5,

10, 30 for seconds, or 1, 2, 5, 10, 30 for minutes, or 1 for hours.

• Secondary display: time unit SEC, MIN, or HOUR as appropriate

for the choice presently selected

• press the ▼ key to decrease the interval selection, stopping at 1

second.

• press the ▲ key to increase the interval selection, stopping at 1

hour.

• press the CONFIG key to step to the next menu item.

• press the ENTER key to exit configuration.

4) Select maximum number of samples

• Main display: number of samples presently selected

• Secondary display: SAMPL

• press the ▼ key to decrease the number, stopping at 1.

• press the ▲ key to increase the number, stopping at the

maximum number of records 8500.

• press the CONFIG key to step to the next menu item.

• press the ENTER key to exit configuration.

5) Erase all logged data and set to Demand mode

• Main display: CLEAr

• Secondary display: DATA

• the ▼ and ▲ keys are ignored.

• press the ENTER key to step to the next menu item and confirm

erasure.

• press the CONFIG to exit configuration without erasing the data.

8