Memory function, 1 how to save the data – KYORITSU 6016 User Manual
Page 50
47
16. MEMORY FUNCTION
Measured result at each function can be saved in the memory of the instrument.
(MAX : 1000)
16.1 How to save the data
Save the result according to following sequence.
(1) Measured result.
(2) Press
to enter into MEMORY MODE.
(3) Press
to enter into SAVE MODE.
(4) Make setting for following items.
1. CIRCUIT No
2. BOARD No
3. SITE No
4. DATA No
Press the SELECT Button to choose the parameter to change.
CIRCUIT No → BOARD No → SITE No → DATA No
Use the UP or DOWN Button and change settings.
Keep the UP/DOWN Key
pressed down to alter the
number quickly.
250V
MΩ
18.52
Fig.48-1
Fig.48-2
Fig.48-3
ALL DELETE
MEMORY MODE
ESC:MEM BUTTON
DELETE
RECALL
SAVE
OK
SAVE MODE
DOWN
UP
SITE:
01
DATA No.001
BOARD:
CIRCUIT:
01
01
SELECT
See also other documents in the category KYORITSU Tools:
- 1009 (13 pages)
- 1011 (12 pages)
- 1012 (12 pages)
- 1018 (1 page)
- 1030 (1 page)
- 1061 (58 pages)
- 1051 (40 pages)
- 1109S (36 pages)
- 1110 (1 page)
- 2000 (1 page)
- 2002PA (27 pages)
- 2007A (1 page)
- 2009R (33 pages)
- 2012R (2 pages)
- 2017 (1 page)
- 2031 (1 page)
- 2033 (9 pages)
- 2040 (2 pages)
- 2046R (2 pages)
- 2200 (3 pages)
- 2210R (1 page)
- 2300R (1 page)
- 2413F (24 pages)
- 2413R (24 pages)
- 2431 (2 pages)
- 2432 (1 page)
- 2433R (1 page)
- 2434 (10 pages)
- 2500 (1 page)
- 2608A (2 pages)
- 3005A (24 pages)
- 3021 (24 pages)
- 3121A (12 pages)
- 3126 (28 pages)
- 3127 (60 pages)
- 3128 (88 pages)
- 3131A (20 pages)
- 3132A (20 pages)
- 3161A (24 pages)
- 3321A (24 pages)
- 4105A (10 pages)
- 4106 (48 pages)
- 4116A (20 pages)
- 4140 (32 pages)
- 4200 (24 pages)