beautypg.com

2 principles of uc measurement, 3 operating instructions for rcd, Fig.37 – KYORITSU 6016 User Manual

Page 42: Fig.38

background image

39

Practical example of RCD test with distribution leads.

10.2 Principles of Uc Measurement
Ground being imperfect in the Fig35, when R exists, when a fault current flows to R,
electric potential occurs. There is a possibility the person contacting in this imperfect
ground, it calls the voltage, which it occurs in the human body of this time, called Uc.
When with the Uc Test letting flow IΔN to the RCD, the Uc is calculated.
Uc voltage is calculated based on the Rated Residual Current (I△N) with the
impedance measured.

10.3 Operating Instructions for RCD
10.3.1 Initial Checks: to be carried out before any testing;
1. Preparation
Always inspect your test instrument and lead accessories for abnormality or damage:
If abnormal conditions exist DO NOT PROCEED WITH TESTING. Have the instrument
checked by your distributor.

Operation of Function Switch

F1

Measurement mode setting
(X1/2, X1, X5, Ramp, Auto, Uc)

F2

IΔn setting

F3

RCD Type setting ( , , , )

F4

PHASE setting (0° ,180° )

PE

×1

ms

30 mA

230V

L-PE

UL50V

PHASE : 0°

50.0Hz

L-PE

L-N

!

Fig.37

×1 / 2

ms

30 mA

L-PE

UL50

V

PHASE : 0°

230V

50.0Hz

L-PE

L-N

!

Fig.38