beautypg.com

KYORITSU 6016 User Manual

Page 12

background image

9

PSC (L-N/L-L) / PFC (L-PE)

Function

Rated

Voltage

Nominal Test Current

at 0Ω External Loop:

Magnitude/Duration(*5)

Range

Accuracy

PSC

100~500V

50/60Hz

6A/20ms

2000A/20kA

Auto-Ranging

PSC/PFC

accuracy is

derived from

measured loop

impedance

specification and

measured voltage

specification

PFC

100~260V

50/60Hz

6A/20ms

2A/20ms

15mA/500ms

PFC

(ATT)

100~260V

50/60Hz

L-N: 6A/60ms

N-PE: 10mA/approx. 5s

*5: at 230V

RCD

Function

Rated Voltage

Accuracy

Trip Current

Trip Time

AC Type

A Type

X1/2

230V+10%-15%

50/60Hz

-8%~-2%

-10%~0%

±(1%rdg+3dgt)

X1

+2%~+8%

0%~+10%

X5

+2%~+8%

0%~+10%

Ramp(◢)

±4%

± 10%

Auto

Depending on the accuracy at each function.

Measurement sequence:

X1/2 0°→X1/2 180°→X1 0°→X1 180°→X5 0°→X5 180°

Measurements with x5 are not carried out for RCDs with nominal

current of 100mA or more.

RCD(Uc)

Function

Rated Voltage

Range

Test Current

Accuracy

UC

230V+10%-15%

50/60Hz

100.0V

1/2I⊿n

(max150mA)

+5% ~ +15%rdg

±8dgt