KYORITSU 1062 User Manual
Page 53
111
l
Calibration of Capacitor Range
Before start calibration of the Capacitor range,
turn the function switch back to the OFF position.
1) Turn the function switch from the OFF position to the
(Capacitor) position while
pressing the SELECT and RANGE keys at the same time.
The display shows the “ CAL ” symbol then the “ PASS ” symbol.
2) Press the SELECT key. (The display shows the “ - ” symbol.)
3) Press the HOLD key twice. (The display shows the “ - - - ” symbol.)
4) Press the RANGE key. (The display shows the “ nF ” symbol.)
5) Connect the instrument to the calibrator with the test leads.
6) Set the calibrator to Input 1 value as an input to the instrument.
7) Press the HOLD key to confirm.
8) Set the calibrator to Input 2 value as an input to the instrument.
9) Press the HOLD key to confirm.
10) Carry out calibration of other ranges by repeating steps 6) to 9).
11) To quit calibration, turn the function switch back to the OFF position.
Range
Input 1
Input 2
Unit
5nF
0.500
5.000
nF
50nF
5.00
50.00
nF
500nF
50.0
500.0
nF
5μF
0.500
5.000
μF
50μF
5.00
50.00
μF
500μF
50.0
500.0
μF
5mF
0.500
5.000
mF
50mF
5.00
40.00
mF